Cementing a polyethylene cup into a well fixed acetabular metal-on-metal resurfacing component? An experimental
Matthias C M Klotz1, Sebastian Jäger1, Jan-Philippe Kretzer1
1Department of Orthopaedics, Trauma Surgery and Spinal Cord Injury, Heidelberg University Hospital, Heidelberg - Germany.
Introduction:
Adverse reactions to metal debris often indicate revision surgery in metal-on-metal (MoM) hip arthroplasty and an exchange of the MoM bearing into either a metal on polyethylene or a ceramic-on-polyethylene articulation. At the moment the removal of the entire implant system is the most reasonable method. In order to avoid bone loss caused by the removal of a well-fixed acetabular component, the purpose of this study was to measure the stability of a cemented polyethylene (PE) cup in an acetabular hip resurfacing component and to examine if such a method could be suitable for clinical use.
Methods:
PE cups were cemented into 2 different hip resurfacing components and biomechanical tests were applied to measure failure torques under lever out and rotational load.
Results:
In all cases failure of the interface between the resurfacing components and the cement layer occurred at a very low load (0.14 Nm-61.50 Nm).
Discussion:
The early failure occurred due to lacking interdigitation of cement and the polished metal surface. Thus we warn against cementing a PE cup into acetabular hip resurfacing components for clinical use.