Related Experiment Video
Updated: Mar 7, 2026

Laryngeal Mask Airway LMA Placement in a Neonatal Patient Simulator Using a Non-Inflatable Supraglottic Airway SGA
Published on: July 14, 2023
Determination of insertion depth of flexible laryngeal mask airway in pediatric population-A prospective
Ji-Hyun Lee1, Hye-Won Oh1, In-Kyung Song1
1Department of Anesthesiology and Pain Medicine, Seoul National University Hospital, #101 Daehak-ro, Jongno-gu, Seoul 110-744, Republic of Korea.
Study Objective:
The purpose of this study was to determine the ideal insertion depth of the flexible laryngeal mask airway (FLMA) by elucidating the relationships between insertion depth and patient's age, body weight, height, and other parameters. We also evaluated an insertion technique that uses the change in intracuff pressure for proper positioning of the FLMA in cases where it is difficult to sense resistance during FLMA insertion.
Design:
This study was a prospective observational study.
Setting:
Participants were recruited from the Seoul National University Children's Hospital.
Patients:
We enrolled 154 children aged ≤15 years with an American Society of Anesthesiologists physical status of I or II who were scheduled for ophthalmic surgery of <2 hours duration under general anesthesia.
Interventions:
After induction of general anesthesia, FLMA insertion was guided by the change in intracuff pressure, measured using a manometer. The FLMA position was assessed using a fiberoptic bronchoscope.
Measurements:
The FLMA insertion depth was measured at the end of each surgical procedure. A multiple linear regression model was then created using age, height, weight, nasal-tragus length, and sternal length.
Main Results:
The FLMA was successfully inserted in the first attempt in 134 patients using continuous monitoring of intracuff pressure. Using multiple linear regression analysis and the Durbin-Watson test, we found that insertion depth was best predicted by height and weight (r2=0.777), and the resulting formula was as follows: insertion depth of FLMA (cm)=7.0+0.04×height (cm)+0.05 ×weight (kg).
Conclusions:
The FLMA insertion depth can be calculated using height and weight. Continuous monitoring of intracuff pressure during FLMA insertion is a useful alternative insertion method in cases where resistance is difficult to sense.
Related Concept Videos
Cardiopulmonary Resuscitation II: ACLS Airway Management
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Suctioning the Nasopharyngeal Airway
Equipment Required
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...