Related Experiment Video
Updated: Mar 7, 2026

Atomic Absorbance Spectroscopy to Measure Intracellular Zinc Pools in Mammalian Cells
Published on: May 16, 2019
Measuring zinc on the Roche cobas c502 analyzer-Validation, comparison, and pre-analytic aspects
Frank Bernhard Kraus1,2, Beatrice Ludwig-Kraus1
1Zentrallabor, Department of Laboratory Medicine, University Hospital Halle, Halle (Saale), Germany.
Objectives:
Characterization of an in vitro diagnostic zinc assay (LT-SYS) on a Roche cobas c502 analyzer and evaluation of the influence of pre-analytic factors on zinc concentration measurements.
Design And Methods:
Imprecision, bias, linearity, limit of blank (LoB), and limit of detection (LoD) were established and method comparisons were performed based on the respective CLSI guidelines. The influence of time elapsed until analysis, the usage of a pneumatic tube delivery system (PTDS) and of special trace element sample tubes was evaluated as well.
Results:
Estimates of imprecision ranged from 0.9% to 5.0% and bias was low with 1.3% and 1.5% deviation from target value. Linearity was met for the measuring range of 1.15-34.7 μmol/L (7.51-226.9 μg/dL), LoB and LoD were 0.17 μmol/L (1.11 μg/dL) and 0.73 μmol/L (4.77 μg/dL) respectively. The method comparison revealed an average deviation of 8.44% (y=0.542+1.036x). Plasma samples had 7.3% higher zinc values than serum samples on the average. Zinc values of uncentrifuged serum and plasma samples increased 20% in 8 hours, while after centrifugation no significant increase could be detected. Usage of PTDS increased zinc values by 17% and usage of trace element sample tubes showed no advantage over normal ones.
Conclusions:
The LT-SYS zinc assay showed a fully acceptable performance with good degrees of imprecision and bias, no deviation from linearity and both a very low LoB and LoD. Samples for zinc analysis should be centrifuged timely and transport over PTDS should be avoided.