Purpose:
Innovations in intraoperative imaging lead to major changes in orthopaedic surgery. In our setting, a 3D flat-panel c-arm (Artis zeego) is mounted on a robotic arm offering an integrated aiming tool (Syngo iGuide). Our aim was to investigate the feasibility of Syngo iGuide for pedicle screw placement in comparison with fluoroscopic screw implantation.
Methods:
In 10 lumbar models, 100 screws were implanted. In 5 models, a standard fluoroscopic technique was used. Syngo iGuide was used in all other models. Afterwards, CT-scans were performed and screw accuracy was investigated.
Results:
The procedure time for the new technique was significantly longer in comparison with the standard technique. The post-operative CT showed the same accuracy in both groups.
Conclusions:
Syngo iGuide proofed feasible for percutaneous implantation of pedicle screws in anatomic models. Syngo iGuide can be a help for screw implantation in difficult anatomic regions without the need of an additional navigation system.