Pitfalls of invasive blood pressure monitoring using the caudal ventral artery in rats
Hiroki Ohta1,2, Takao Ohki2, Yuji Kanaoka2
1Division of Regenerative Medicine, Research Center for Medical Sciences, the Jikei University School of Medicine, 3-25-8 Nishi-shinbashi, Minato-ku, Tokyo 105-8461, Japan.
Abstract:
During rodent experiments, the caudal ventral artery (CVA) is useful for blood pressure (BP) measurement. However, CVA measurements may not reflect the true BP. This study was performed to verify the site-specific accuracy of invasive arterial BP monitoring during surgery in rats. Invasive arterial BP was simultaneously measured in rats via the CVA and the common carotid artery (CCA). The BP values were analysed while the rats were subjected to cooling of the head or tail. Additionally, the rats underwent digital subtraction angiography and histological examination of these arteries. The pressure difference was more significant in the tail cooling group than in the head cooling group. Digital subtraction angiography revealed that angiospasms occurred more frequently in the CVA than in the CCA upon cooling. This phenomenon was supported by histological analysis, which showed that the tunica media area was significantly larger in the CVA than in the CCA. CVA pressure is susceptible to environmental changes and may not accurately reflect the true BP without a strictly controlled laboratory environment. Therefore, understanding the pitfalls of this method is necessary to avoid cooling of the tail during BP measurement.
More Related Videos
09:38Femoral Arterial and Venous Catheterization for Blood Sampling, Drug Administration and Conscious Blood Pressure and Heart Rate Measurements
Published on: January 24, 2012
09:37Invasive Hemodynamic Characterization of the Portal-hypertensive Syndrome in Cirrhotic Rats
Published on: August 1, 2018