Septic transfusion case caused by a platelet pool with visible clotting due to contamination with Staphylococcus

Maria Loza-Correa1, Yuntong Kou1, Mariam Taha1

  • 1Canadian Blood Services.

Transfusion
|February 17, 2017
PubMed
Abstract

Insights

Staphylococcus aureus contamination in platelet concentrates (PCs) poses a significant transfusion risk. A false-negative screening led to a severe transfusion reaction, highlighting the need for improved detection methods and visual inspection of blood products.

Area of Science:

  • Transfusion Medicine
  • Microbiology
  • Hematology

Background:

  • Platelet concentrate (PC) contamination with Staphylococcus aureus is a critical transfusion safety concern.
  • Routine screening for bacterial contamination in PCs can yield false-negative results.

Observation:

  • A case of transfusion reaction occurred in an elderly acute myeloid leukemia patient receiving a 4-day-old PC.
  • The PC was found to be contaminated with S. aureus, which also infected the patient.
  • S. aureus formed biofilms and produced enterotoxin-like U in the transfused PCs.

Findings:

  • The implicated PC tested negative in routine bacterial screening.
  • The patient developed septic symptoms following transfusion.
  • Identical S. aureus strains were isolated from the PC and the patient, confirmed by genome sequencing.

Implications:

  • Point-of-care testing could prevent such false-negative transfusion sepsis cases.
  • Visual inspection of blood components is crucial for identifying contamination.
  • The increasing incidence of S. aureus as a PC contaminant has serious clinical consequences.