Highly Cross-linked Polyethylene Liner Dissociation from a Cement-less Modular Acetabular Shell: Two Case Reports
Shunsuke Kawano1, Motoki Sonohata1, Masaru Kitajima1
1Department of Orthopaedic Surgery, Faculty of Medicine, Saga University, 5-1-1 Nabeshima, Saga 849-8501, Japan.
The Open Orthopaedics Journal
|February 21, 2017
Summary
A rare complication in total hip arthroplasty (THA), polyethylene liner dissociation occurred in 0.04% of cases. This study suggests a potential issue with the push-in liner fixation design.
Area of Science:
- Orthopedic Surgery
- Biomaterials Science
- Medical Device Engineering
Background:
- Total hip arthroplasty (THA) is a common procedure for hip joint issues.
- Polyethylene liner dissociation from cementless acetabular sockets is a rare but serious complication.
- Understanding the failure mechanisms of THA components is crucial for improving patient outcomes.
Purpose of the Study:
- To report and analyze two cases of polyethylene liner dissociation from a cementless acetabular socket.
- To investigate the potential causes of liner dissociation in the context of specific implant models.
- To contribute to the understanding of rare complications in THA to inform future implant design.
Main Methods:
- Retrospective review of two cases of polyethylene liner dissociation following THA.
- Analysis of patient demographics, surgical history, and implant details (26 mm femoral head, CPE liner, acetabular inclination).
- Examination of the explanted polyethylene liners for signs of damage, including partial rim damage and scratches.
Main Results:
- Two cases (0.04%) of cross-linked polyethylene liner dissociation from AMS-HA shells occurred approximately 10 years post-THA.
- No implant loosening was observed in either case.
- Explanted liners showed partial damage to the elevated rim and scratches on the back, suggesting a mechanical failure related to the fixation mechanism.
Conclusions:
- The observed liner dissociation is likely attributed to a design flaw in the push-in type liner fixation of the specific acetabular socket model.
- This complication, though rare, highlights the importance of evaluating liner fixation integrity in cementless THA.
- Further investigation into liner locking mechanisms is recommended to prevent such rare adverse events.