Related Experiment Video
Updated: Mar 7, 2026

Closed Chest Biventricular Pressure-Volume Loop Recordings with Admittance Catheters in a Porcine Model
Published on: May 18, 2021
Concomitant pulsatile and continuous flow VAD in biventricular and univentricular physiology: a comparison study with
Arianna Di Molfetta1, Gianfranco Ferrari2, Roberta Iacobelli1
1Department of Pediatric Cardiology and Cardiac Surgery, Bambino Gesù Pediatric Hospital, Rome - Italy.
Introduction:
To develop and test a lumped parameter model to simulate and compare the effects of the simultaneous use of continuous flow (CF) and pulsatile flow (PF) ventricular assist devices (VADs) to assist biventricular circulation vs. single ventricle circulation in pediatrics.
Methods:
Baseline data of 5 patients with biventricular circulation eligible for LVAD and of 5 patients with Fontan physiology were retrospectively collected and used to simulate patient baselines. Then, for each patient the following simulations were performed: (a) CF VAD to assist the left ventricle (single ventricle) + a PF VAD to assist the right ventricle (cavo-pulmonary connection) (LCF + RPF); (b) PF VAD to assist the left ventricle (single ventricle) + a CF VAD to assist the right ventricle (cavo-pulmonary connection) (RCF + LPF).
Results:
In biventricular circulation, the following results were found: cardiac output (17% RCF + LPF, 21% LCF + RPF), artero-ventricular coupling (-36% for the left ventricle and -21.6% for the right ventricle), pulsatility index (+6.4% RCF + LPF, p = 0.02; -8.5% LCF + RPF, p = 0.00009). Right (left) atrial pressure and right (left) ventricular volumes are decreased by the RCF + LPF (by RPF + LCF). Pulmonary arterial pressure decreases in the LCF + RPF configuration. In Fontan physiology: cardiac output (LCF + RPF 35% vs. 8% in RCF + LPF), ventricular preload (+4% RCF + LPF, -10% LCF + RPF), Fontan conduit pressure (-5% RCF + LPF, +7% LCF + RPF), artero-ventricular coupling (-14% RCF + LPF vs. -41% LCF + RPF) and pulsatility (+13% RCF + LPF, - 8% LCF + RPF).
Conclusions:
A numerical model supports clinicians in defining and innovating the VAD implantation strategy to maximize the hemodynamic benefits. Results suggest that the hemodynamic benefits are maximized by the LCF + RPF configuration.
Related Concept Videos
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Aortic Regurgitation I: Introduction
Mitral Regurgitation I: Introduction
Autoregulation of Blood Flow
Chemical Signaling in Autoregulation
Chemical signaling operates at the precapillary sphincter level, inciting either contraction or relaxation....
Imbalances in Cardiac Output
CHF can occur due to the failure of either side of the heart. Left-side failure leads to pulmonary congestion—the right side continues to send...