Related Experiment Video
Updated: Mar 7, 2026

Fully Endoscopic Mitral Valve Repair with Percutaneous Cannulation of Groin Vessels
Published on: May 26, 2023
Mechanical recanalization for clot occlusion of venous access ports: experimental study using ports with clot
Myung Gyu Song1, Tae-Seok Seo1, Baek-Hui Kim2
1Department of Radiology, Korea University Guro Hospital, Korea University College of Medicine, Seoul - Korea.
Purpose:
To test the hypothesis that mechanical injection of saline is safe and effective in restoring patency of totally implantable venous access ports (TIVAPs) with clot occlusion. We devised an experimental port model for the evaluation of mechanical TIVAP recanalization prior to its clinical application.
Materials And Methods:
The clot TIVAP occlusion model was constructed by filling the catheter with swine blood and incubating it at 37.5°C. The model was incubated for different lengths of time ranging from 1 day to 7 days. Each incubation time point included 20 ports. Total catheter occlusion of the TIVAPs was assessed with a 10-mL saline syringe equipped with a non-coring needle. Occlusion was defined as no passage of saline through the catheter when it was aspirated and infused gently with the 10-mL saline syringe. Pressure was evaluated during recanalization with an indeflator. Histological examination was performed on the clot obtained during recanalization.
Results:
Among the 140 total experimental ports, 65 occlusions (46.4%) were detected. Of these 65 occlusions, 56 (86.1%) were recanalized by mechanical saline pressure via the indeflator. The indeflator pressure ranged from 29 pound per square-inch (psi) to 265 psi at mechanical catheter recanalization (mean: 110 psi). Histologically, all specimens from the model ports exhibited a similar appearance; specifically, erythrocytes, cells, and fibrin were evenly scattered throughout the clot.
Conclusions:
Our data indicate that it is feasible to generate a TIVAP clot occlusion model with swine blood. Moreover, mechanical recanalization was suitable for resolving occluded catheters without thrombolytic agents.
More Related Videos
05:51Computed Tomography (CT) Guided Implantation of a Totally Implantable Venous Access Port (TIVAP) through Subclavian Vein
Published on: January 13, 2026
08:05Occlusion of the Great and Small Saphenous Vein Using Copolymeric Glue Based on N-Butyl Cyanoacrylate and Methacryloxy Sulfolane
Published on: December 9, 2022
Related Concept Videos
Venous Thrombosis III: Interprofessional Care
Continuous Renal Replacement Therapy
Venous Thrombosis II: Clinical Manifestations and Diagnostic Studies