Effect of Sonication Duration and Power on Ablation Depth During MR-Guided Focused Ultrasound of Bone
Matthew D Bucknor1, Eugene Ozhinsky1, Rutwik Shah1
1Department of Radiology and Biomedical Imaging, University of California San Francisco, San Francisco, California, USA.
Purpose:
To evaluate the effect of differences in sonication duration and power on the size of postcontrast ablation zone following magnetic resonance-guided focused ultrasound (MRgFUS) of bone in a swine femoral bone model.
Materials And Methods:
Experimental procedures received approval from the Institutional Committee on Animal Research. MRgFUS was used to create two thermal lesions in the left femur of six pigs. Each target was subjected to six sonications. 400J of energy was used for each sonication. However, the distal target received the standard sonication duration of 20 seconds (20W), while the proximal target received a longer sonication duration of 40 seconds (10W). MRgFUS lesions were imaged with fat-saturated spoiled gradient echo sequence at 3.0T MRI 10 minutes following the administration of contrast. Maximum three-plane dimensions of the hypoenhanced ablation area were measured.
Results:
Postcontrast MR images demonstrated ovoid regions of hypoenhancement at each target. The average depth of ablation was significantly greater for the shorter high-power sonications (7.3 mm), compared to the longer lower-power sonications (4.5 mm), P = 0.026. The craniocaudal dimension was also greater for the shorter ablations 26.7 mm compared to the longer sonications 21.0 mm, P = 0.006.
Conclusion:
Contrary to anecdotal clinical experience, this preclinical model suggests that during MRgFUS of bone, standard duration, higher-power sonications resulted in deeper ablation volumes compared to long duration, lower-power sonications. These results suggest that to achieve deeper ablations, if longer sonications are used, then the power should be relatively maintained, for a net energy increase.
Level Of Evidence:
1 Technical Efficacy: Stage 5 J. Magn. Reson. Imaging 2017;46:1418-1422.
More Related Videos
05:54MRI-guided Focused Ultrasound Thalamotomy for Patients with Medically-refractory Essential Tremor
Published on: December 13, 2017
07:38Real-time Monitoring of High Intensity Focused Ultrasound HIFU Ablation of In Vitro Canine Livers Using Harmonic Motion Imaging for Focused Ultrasound HMIFU
Published on: November 3, 2015
Related Concept Videos
Uniform Depth Channel Flow: Problem Solving
Uniform Depth Channel Flow
Power
Depth Perception and Spatial Vision
Instantaneous Power
Complex Power
Complex power is defined as the multiplication of the voltage and the complex conjugate of the current. The magnitude of this power, known as apparent power, is measured in volt-amperes (VA). Notably, the angle of the complex power equates to the...