Theodore F Saad1, Henry L Weiner2
1Section of Renal and Hypertensive Diseases, Christiana Care Health System, Newark, Delaware.
Insights
Patients with end-stage renal disease requiring cardiovascular implantable electronic devices and hemodialysis catheters face high risks. Coordinated care and alternative strategies can minimize complications and improve outcomes.
Area of Science:
- Nephrology
- Cardiology
- Medical Devices
Background:
- End-stage renal disease (ESRD) often presents with cardiac comorbidities requiring cardiovascular implantable electronic devices (CIEDs) like pacemakers or implantable cardioverter-defibrillators.
- In the US, up to 10.5% of chronic hemodialysis (HD) patients have CIEDs, and 18% use venous HD catheters for vascular access.
- The concurrent use of HD catheters and CIEDs poses significant risks, including central venous stenosis and infectious complications.
Abstract:
End-stage renal disease is frequently accompanied by cardiac comorbidity that warrants treatment with a cardiovascular implantable electronic device (permanent pacemaker or implantable cardioverter-defibrillator). In the United States, chronic hemodialysis (HD) population, cardiac implantable devices are present in up to 10.5% of patients; a venous HD catheter is utilized for blood access in 18% of prevalent patients. The concomitant presence of a venous HD catheter and cardiovascular implantable device creates a high-risk circumstance, with potential for causing symptomatic central venous stenosis, and for developing complicated endovascular infection. This dangerous combination may be avoided for many patients by utilizing nondialysis methods for management of advanced chronic kidney disease, initiating dialysis without venous catheter access, or managing cardiac rhythm disorders without use of transvenous cardiac implantable electronic devices. In those situations where the combination of a venous HD catheter and cardiac implantable device is unavoidable, there are strategies to minimize duration of venous catheter access, and to reduce risks for infectious complications. It is essential for nephrologists and cardiologists to understand the indications, alternatives, and risks involved with venous HD access and cardiac implantable devices. Coordinated management of renal disease and cardiac rhythm disorders has potential to minimize risks, improve outcomes, and substantially reduce the cost of care.