Evaluation of interarm blood pressure differences using the Microlife WatchBP Office in a clinical setting
Christoffer Krogager1, Esben Laugesen, Niklas B Rossen
1aDepartment of Endocrinology and Internal Medicine, Aarhus University Hospital bDepartment of Clinical Medicine cSection for Biostatistics, Department of Public Health, Aarhus University, Aarhus dDiagnostic Centre, Silkeborg Regional Hospital, Silkeborg eThe Danish Diabetes Academy, Odense University Hospital, Odense, Denmark.
Objective:
The aim of this study was to evaluate the usefulness of Microlife WatchBP Office and the effect of increasing the number of measurements in the clinical evaluation of systolic interarm difference (IAD).
Patients And Methods:
Office blood pressure was measured simultaneously on both arms in 339 patients (85% diabetic) using the Microlife WatchBP Office, a fully automatic, oscillometric device. The patients included were all scheduled for ambulatory blood pressure measurement at the outpatient clinic of endocrinology at Silkeborg Regional Hospital, Denmark. Two successive sets of three individual measurements were made. A statistical analysis of variance was carried out on the measurements.
Results:
In the first set of measurements, the mean IAD was -0.3 mmHg and the prevalence of IAD greater than or equal to 10 mmHg was 9.1%. Only 7.6% of the patients with an IAD less than 10 mmHg in the first set of measurements had an IAD greater than or equal to 10 mmHg in the second set of measurements. The 95% limits of agreement for the mean IAD for a single set of three measurements were ±13.16 mmHg. The probability of detecting an IAD more than 10 mmHg only increased slightly with an increasing number of measurements.
Conclusion:
A single set of triplicate measurements using Microlife WatchBP is an acceptable method for evaluating IAD as more measurements do little to improve the probability of detecting an IAD more than 10 mmHg because of high intraindividual variation.
More Related Videos
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
05:51Measuring the Carotid to Femoral Pulse Wave Velocity Cf-PWV to Evaluate Arterial Stiffness
Published on: May 3, 2018
Related Concept Videos
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Pre-Procedural Guidelines for Assessing Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Measurement of Blood Pressure