Related Experiment Video
Updated: Mar 7, 2026

Author Spotlight: Microbial Control and Monitoring Strategies for Cleanroom Environments and Cellular Therapies
Published on: March 17, 2023
An approach for estimating measurement uncertainty in medical laboratories using data from long-term quality control
Background:
The present study was prompted by the ISO 15189 requirements that medical laboratories should estimate measurement uncertainty (MU).
Methods:
The method used to estimate MU included the: a) identification of quantitative tests, b) classification of tests in relation to their clinical purpose, and c) identification of criteria to estimate the different MU components. Imprecision was estimated using long-term internal quality control (IQC) results of the year 2016, while external quality assessment schemes (EQAs) results obtained in the period 2015-2016 were used to estimate bias and bias uncertainty.
Results:
A total of 263 measurement procedures (MPs) were analyzed. On the basis of test purpose, in 51 MPs imprecision only was used to estimate MU; in the remaining MPs, the bias component was not estimable for 22 MPs because EQAs results did not provide reliable statistics. For a total of 28 MPs, two or more MU values were calculated on the basis of analyte concentration levels. Overall, results showed that uncertainty of bias is a minor factor contributing to MU, the bias component being the most relevant contributor to all the studied sample matrices.
Conclusions:
The model chosen for MU estimation allowed us to derive a standardized approach for bias calculation, with respect to the fitness-for-purpose of test results. Measurement uncertainty estimation could readily be implemented in medical laboratories as a useful tool in monitoring the analytical quality of test results since they are calculated using a combination of both the long-term imprecision IQC results and bias, on the basis of EQAs results.
Related Concept Videos
Quality Assurance
Quality Control
Quality control helps track data, visualize trends, and identify variations, making it easier to detect deviations that may affect the accuracy of an analysis. One way to do this is by generating a quality control chart, which...
Uncertainty in Measurement: Accuracy and Precision
Data Validation
Key parameters for method validation include:
Uncertainty: Overview
Uncertainty in Measurement: Reading Instruments