Quality assurance guidelines for superficial hyperthermia clinical trials : II. Technical requirements for heating
Hana Dobšíček Trefná1, Johannes Crezee2, Manfred Schmidt3
1Signals and Systems, Chalmers University of Technology, Gothenburg, Sweden. hanatre@chalmers.se.
Abstract:
Quality assurance (QA) guidelines are essential to provide uniform execution of clinical trials with uniform quality hyperthermia treatments. This document outlines the requirements for appropriate QA of all current superficial heating equipment including electromagnetic (radiative and capacitive), ultrasound, and infrared heating techniques. Detailed instructions are provided how to characterize and document the performance of these hyperthermia applicators in order to apply reproducible hyperthermia treatments of uniform high quality. Earlier documents used specific absorption rate (SAR) to define and characterize applicator performance. In these QA guidelines, temperature rise is the leading parameter for characterization of applicator performance. The intention of this approach is that characterization can be achieved with affordable equipment and easy-to-implement procedures. These characteristics are essential to establish for each individual applicator the specific maximum size and depth of tumors that can be heated adequately. The guidelines in this document are supplemented with a second set of guidelines focusing on the clinical application. Both sets of guidelines were developed by the European Society for Hyperthermic Oncology (ESHO) Technical Committee with participation of senior Society of Thermal Medicine (STM) members and members of the Atzelsberg Circle.
Related Concept Videos
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Decreased Body Temperature
Assessing Body Temperature - Axilla
Step 1: Perform hand hygiene and put on clean gloves to maintain infection control and prevent cross-contamination.
Step 2: Prepare the patient by explaining the procedure to ensure understanding and cooperation. Ensure privacy, expose the axilla, and inform the patient that minimal movement is crucial for an accurate reading.
Step 3: Adjust the patient’s clothing to expose only the axilla. It minimizes...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...