Magnetocardiography measurements with 4He vector optically pumped magnetometers at room temperature
S Morales1, M C Corsi1, W Fourcault1
1CEA, LETI, MINATEC Campus, F-38054 Grenoble, France.
None:
In this paper, we present a proof of concept study which demonstrates for the first time the possibility of recording magnetocardiography (MCG) signals with 4He vector optically pumped magnetometers (OPM) operated in a gradiometer mode. Resulting from a compromise between sensitivity, size and operability in a clinical environment, the developed magnetometers are based on the parametric resonance of helium in a zero magnetic field. Sensors are operated at room temperature and provide a tri-axis vector measurement of the magnetic field. Measured sensitivity is around 210 f T (√Hz)-1 in the bandwidth (2 Hz; 300 Hz). MCG signals from a phantom and two healthy subjects are successfully recorded. Human MCG data obtained with the OPMs are compared to reference electrocardiogram recordings: similar heart rates, shapes of the main patterns of the cardiac cycle (P/T waves, QRS complex) and QRS widths are obtained with both techniques.