Radiation Exposure of Patients and Interventional Radiologists during Prostatic Artery Embolization: A Prospective
Gustavo Andrade1, Helen J Khoury2, William J Garzón2
1AngioRad-Interventional Radiology, Ribeirão Preto, Brazil; Faculty of Medicine, Universidade de São Paulo, Ribeirão Preto, Brazil.
Purpose:
To prospectively analyze the radiation exposure of patients and interventional radiologists during prostatic artery embolization (PAE).
Materials And Methods:
Twenty-five consecutive PAE procedures performed with an Artis zee system in a single center by an interventional radiologist were prospectively monitored. The mean age, weight, and prostate volume of the patients were 65.7 year (range, 43-85 y), 71.4 kg (range, 54-88 kg), and 79 cm3 (range, 36-157 cm3), respectively. In addition to Digital Imaging and Communications in Medicine radiation data, direct measures were also obtained. Radiochromic film was used to evaluate peak skin dose (PSD). The radiologist wore a protective apron and a thyroid collar, and a ceiling-suspended screen and a table curtain were used. To estimate the absorbed doses, nine pairs of dosimeters were attached to the operator's body.
Results:
The average fluoroscopy time was 30.9 minutes (range, 15.5-48.3 min). The mean total dose-area product (DAP) was 450.7 Gy·cm2 (range, 248.3-791.73 Gy·cm2) per procedure. Digital subtraction angiography was responsible for 71.5% of the total DAP, followed by fluoroscopy and cone-beam computed tomography. The mean PSD was 2,420.3 mGy (range, 1,390-3,616 mGy). The average effective dose for the interventional radiologist was 17 μSv (range, 4-47 μSv); values for the eyes, hands, and feet were obtained, and were all greater on the left side.
Conclusions:
PAE may lead to high x-ray exposures to patients and interventional radiologists.
Related Concept Videos
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies VII: Vascular Imaging
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Pulmonary Embolism II: Diagnostic Studies and Interprofessional Care