Related Experiment Video
Updated: Mar 6, 2026

One-day Workflow Scheme for Bacterial Pathogen Detection and Antimicrobial Resistance Testing from Blood Cultures
Published on: July 9, 2012
Bacterial Contamination of CT Equipment: Use of ATP Detection and Culture Results to Target Quality Improvement
John Childress1, Debborah Burch1, Cheryl Kucharski1
1Department of Radiology, University of Michigan Health System, 1500 E. Medical Center Dr. B2-A209P, Ann Arbor, MI 48108.
Rationale And Objective:
This study aimed to evaluate the use of an adenosine triphosphate (ATP) monitoring system to minimize surface contamination on inpatient computed tomography (CT) scanners.
Methods:
The bore, table, and wrap of two quaternary care inpatient CT scanners (load/scanner: ~ 30-40 CT examinations/day) were assayed with bacterial cultures and an ATP detection system during six prospective iterative plan-do-check-act improvement cycles from January 6, 2016 to October 12, 2016. Per-cycle sampling was for eight consecutive weekdays. ATP detection was expressed as relative light units (RLUs) through a luciferase reaction, with >350 RLU considered contaminated per manufacturer recommendations. Culture swabs were placed into 6.5% NaCl broth, a Staphylococcus enrichment broth, and incubated aerobically at 37°C for 48 hours. Positive broths were plated to chromogenic Staphylococcus media. Culture rates (Fisher exact test) and RLU values (Mann-Whitney U test) were compared.
Results:
In Cycle 1, both culture results and median RLU values indicated the wrap was the most contaminated item (positive culture rate: 63% [10/16], median RLU interquartile range: 173 [IQR: 56-640]); however, RLU values were not predictive of per-sample culture results (P = .36). Following iterative improvements, RLU values at Cycle 6 were significantly lower than at peak (P = .02-.04) and within manufacturer's recommendations: all samples: 45 (IQR: 16-87), bore: 26 (IQR: 0-51), table: 68 (IQR: 21-89), wrap: 47 (IQR: 38-121).
Conclusion:
The Velcro wrap is the most contaminated item on a CT scanner, and special processes may be needed to ensure adequate cleansing. ATP detection is a crude surrogate for bacterial culture results but benefits from speed, reduced cost, and greater statistical power.