Validity of computed tomography in predicting scaphoid screw prominence: a cadaveric study
Clare E Griffis1, Cara Olsen2, Leon Nesti3
1Captain James A. Lovell Federal Health Care Center, USS Tranquillity BHC 1007, 3001 Greenbay Road, North Chicago, IL, 60064, USA. claregriffis@gmail.com.
Background:
Studies of hardware protrusion into joint spaces following fracture fixation have been performed to address whether or not there is discrepancy between the actual and radiographic appearance of screw prominence. The purpose of our study was to prove that, with respect to the scaphoid, prominence as visualized on CT scan is real and not a result of metal artifact.
Methods:
Forty-two cadaveric wrists were separated into four allotted groups with 21 control specimens and 21 study specimens. All specimens were radiographically screened to exclude those with inherent carpal abnormalities. Acutrak® headless compression screws were placed into all specimens using an open dorsal approach. Cartilage was removed from screw insertion site at the convex surface of the scaphoid proximal pole. Control specimens had 0 mm screw head prominence. The studied specimens had 1, 2, and 3 mm head prominence measured with a digital caliper. Computed tomography, with direct sagittal acquisition and metal suppression technique, was then performed on all specimens following screw placement. Two staff radiologists blinded to the study groups interpreted the images.
Results:
Results revealed that only one of 21 control specimens was interpreted as prominent. Comparatively, in the studied groups, 90% were accurately interpreted as prominent.
Conclusions:
CT provides an accurate assessment of scaphoid screw head prominence. When a screw appears prominent on CT scan, it is likely to be truly prominent without contribution from metallic artifact.
More Related Videos
08:39Longitudinal Micro-Computed Tomography Image Analysis for User-Defined Region of Interest in Critical-Sized Bone Defects
Published on: June 24, 2025
09:34A Method to Estimate Cadaveric Femur Cortical Strains During Fracture Testing Using Digital Image Correlation
Published on: September 14, 2017
Related Concept Videos
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...