Commissioning of beam shaper applicator for conformal intraoperative electron radiotherapy
Nematollah Heidarloo1, Hamid Reza Baghani1, Seyed Mahmoud Reza Aghamiri1
1Radiation Medicine Department, Shahid Beheshti University, Daneshjou St., Velenjak, 1983963113 Tehran, Iran.
Abstract:
Beam shaper is a newly designed device for beam shaping in IOERT. The aim of this study is evaluating the performance of this device for conformal IOERT and preparing it for clinical applications. Dosimetric characteristics for different combinations of energy/field size were determined by ionometric dosimetry, film dosimetry and Monte Carlo simulation. Desirable dosimetric characteristics of beam shaper make it a useful tool for conformal IOERT. Usability of Monte Carlo simulation in any clinical setup of beam shaper was demonstrated.