Related Experiment Video
Updated: Mar 6, 2026

Implantation of Total Artificial Heart in Congenital Heart Disease
Published on: July 18, 2014
First experiences with HeartMate 3 follow-up and adverse events
Jasmin S Hanke1, Günes Dogan1, Sebastian V Rojas1
1Department of Cardiac, Thoracic, Transplantation, and Vascular Surgery, Hannover Medical School, Hannover, Germany.
Background:
The novel HeartMate 3 (HM3) left ventricular assist device (LVAD) received its CE mark in October 2015. It is a new compact LVAD featuring fully magnetically levitated pump, artificial pulse, large pump gaps, and a modular driveline. Here, we present outcomes and adverse events of a single-center cohort 6 months after HM3 implantation.
Methods:
We retrospectively studied a patient cohort of 27 patients who were supported with the HM3 at a single institution. We excluded patients with biventricular assist devices and other types of assist devices as well as LVAD exchange and re-operative procedures.
Results:
Twenty-seven patients were enrolled into the study. Within 1 year after HM3 implantation, 1 patient received a heart transplant and 3 patients died. Thirty-day survival was 88.9% and 6 months 85.2%. No pump thrombosis and no strokes were observed within 6 months. Right heart failure was diagnosed in 1 patient after HM3 implantation (3.7%). No technical complications of the pump were documented. No pump exchanges were necessary.
Conclusions:
The novel LVAD HM3 has already shown good CE mark trial results. Within this first report after the CE mark trial, the 6-month survival after HM3 implantation was 85.2%. The HM3 showed excellent midterm results with 0% stroke and 0% pump thrombosis rates 6 months after implantation.