Related Experiment Video
Updated: Mar 6, 2026

Combined Shuttle-Box Training with Electrophysiological Cortex Recording and Stimulation as a Tool to Study Perception and Learning
Published on: October 22, 2015
Design of new multi-channel electrodes for surface electromyography signals for signal-processing
Abstract:
This paper covers the design aspects of a new multi-channel electrode for the acquisition of surface electromyography signals from a selected muscle. The new multi-channel electrode has 11 pins where the monopolar signals produced will be configured in a software either as Linear array or Laplacian configuration. The design specification of the pre-amplifier ideally was to have a voltage gain of 500 with bandpass filtering of 5 Hz-1 kHz. The final design of the pre-amplifier circuit using an INA 118 instrumentation amplifier was built and tested to give values for voltage gain of 484 with bandpass filtering of 6.8 Hz-1.02 kHz. The software configuration that gives clearer and more defined signals in terms of motor unit action potentials for future signal processing is the Laplacian rather than Linear array.

