Application of Low Tube Voltage 70 kV and Advanced Modeled Iterative Reconstruction in the Third-generation
Jingjuan Liu1, Weidong Pan1, Huadan Xue1
1Department of Radiology,PUMC Hospital,CAMS and PUMC,Beijing 100730,China.
Abstract:
Objective To explore the feasibility of reducing radiation dose of CT colonography (CTC) while preserving image quality using the third-generation dual-source CT at 70 kV tube voltage with advanced modeled iterative reconstruction (ADMIRE). Methods This prospective study consecutively included 53 patients. All patients underwent CTC at 120 kV in supine position with filtered back projection reconstruction and at 70 kV in prone position with ADMIRE with the same scan range. Radiation dose and image quality were compared between these two protocols. Results CT dose index volume,dose-length product,and effective dose of 70 kV were (1.47±0.23)mGy,(68.85±13.82)mGy·cm,and (1.03±0.21)mSv,respectively,which were significantly lower than 120 kV mode [(2.96±0.61)mGy (t=25.356,P=0.000),(136.97±32.77)mGy·cm (t=22.246,P=0.000),and (2.05±0.49)mSv (t=22.293,P=0.000)]. Signal to noise ratio of 70 kV with ADMIRE was 3.60±0.58,which was obviously higher than that of 120 kV as 2.34±0.47 (t=-18.036,P=0.000). Noise of 70 kV with ADMIRE was (11.81±1.93)HU,which was obviously lower than that of 120 kV as (14.39±2.57)HU (t=7.856,P=0.000). The κ value of two radiologists' overall subjective image score was 0.843 (P=0.000). For 2D images,the noise score of 70 kV with ADMIRE was significantly lower (Z=325.000,P=0.000) and the overall image quality was significantly higher (Z=78.000,P=0.001) compared with 120 kV;however,the sharpness had no significant difference. All scores ranged from 4 to 5. For 3D virtual endoscopy,the overall image quality of 70 kV with ADMIRE and 120 kV was not significantly different (Z=4.000,P=0.059),and all scores ranged from 4 to 5. Conclusion CTC at 70 kV tube voltage combined with ADMIRE can remarkably reduce the radiation dose while maintaining imaging quality than 120 kV at third-generation dual-source CT.
More Related Videos
09:49Dual-phase Cone-beam Computed Tomography to See, Reach, and Treat Hepatocellular Carcinoma during Drug-eluting Beads Transarterial Chemo-embolization
Published on: December 2, 2013
06:59Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
Related Concept Videos
Imaging Studies III: Gastrointestinal Motility Studies and Virtual Colonoscopy
Radionuclide Testing
Radionuclide testing is a sophisticated medical technique for assessing gastrointestinal motility. It focuses on gastric emptying and colonic transit time. Radioactive markers track the movement of food through the digestive system, providing insights into gastrointestinal disorders.
In gastric emptying studies, a meal's liquid and...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Endoscopic Procedures II: Colonoscopy
Radiological Investigation I: X-ray and CT