Energy and field size dependence of a silicon diode designed for small-field dosimetry
Mehran Yarahmadi1, Sonja Wegener2, Otto A Sauer2
1Department of Medical Physics, Faculty of Medicine, Kurdistan University of Medical Sciences, Sanandaj, Iran.
Purpose:
To investigate the energy dependence/spectral sensitivity of silicon diodes designed for small-field dosimetry and obtain response factors (RFs) for arbitrary photon spectra using Monte Carlo (MC) simulations.
Methods:
The EGSnrc user-code DOSRZnrc was used to calculate the dose deposition in water and in the active volume of a stereotactic diode field detector (SFD). Then, the RFs of the SFD were calculated for several circular field sizes and energies at 5 cm depth in water. Several low-energy photon spectra (mean energy 55 to 200 keV), as well as Co-60 radiation (mean energy 1.25 MeV) and a 6 MV Elekta Synergy beam (mean energy 2.9 MeV), in 10 × 10 cm2 field size were used to validate the MC calculations, using a simple beam model. The RFs of the SFD detector for a 6 MV Elekta Synergy linac photon beam in different field sizes were calculated. These were also measured with EBT3 Gafchromic film and the SFD detector.
Results:
For the reference field size, differences between measured and calculated RFs were less than 5% at mean energies below 1 MeV and less than 1% at energies above 1 MeV. The calculated RFs for a 6 MV Elekta Synergy linac photon beam as a function of different field sizes showed a good agreement between the measurements and previously reported results. This agreement was within 2% for all considered field sizes.
Conclusion:
While at high photon energies, the change of response of the SFD is marginal, whereas it is extreme at low energies. Therefore, it is desirable to benchmark response calculations also in the low-energy domain. Our results, with a simple beam model and geometry, indicate that a validation of the simulations by experimental results is achievable. The present work provides a comprehensive table that can be used to calculate SFD detector response factors depending on both, field size and photon energy.
Related Concept Videos
Small-signal Diode Model
Biological Effects of Radiation
Diode: Forward bias
The behavior of a diode in forward bias...
Diamagnetic Shielding of Nuclei: Local Diamagnetic Current
Electric Field of a Charged Disk
The system's symmetry is in the cylindrical directions across the plane of the charge. As a result, the electric fields created by various surface charge elements nullify each other in the direction parallel to the surface. Thereby, the resulting electric field is perpendicular to the plane. Since the disk is...
Determining Electric Field From Electric Potential
In general, regardless of whether the electric field is uniform, it points in the direction of decreasing potential because the force on a positive...