Sterols in infant formulas: validation of a gas chromatographic method
Islam J A Hamdan1, Lorena Claumarchirant1, Guadalupe Garcia-Llatas1
1a Nutrition and Food Science Area, Faculty of Pharmacy , University of Valencia , Valencia , Spain.
Abstract:
Sterols are components present in the fat fraction of infant formulas (IFs). Their characterization is therefore of interest, though there are no official reference methods for their analysis in these matrices.
Aim:
To validate a gas chromatographic method with flame ionization detection for the determination of animal (cholesterol and desmosterol) and plant sterols (brassicasterol, campesterol, stigmasterol, β-sitosterol and sitostanol) found in IFs. All correlation coefficients obtained for the calibration curves of sterols studied were >0.99. Limits of detection (<1 μg/100 mL) and quantification (<4 μg/100 mL) are suitable for sterols determination in IFs. The within-assay precision ranged from 1.6% to 8.8%, while the between-assay precision was <10% for most of sterols. Accuracy was satisfactory and was calculated by recovery assays (ranging 93-108%). The analytical parameters obtained showed the suitability of the proposed method for the determination of sterols in IFs.
More Related Videos
Related Concept Videos
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a...
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
Chromatographic Methods: Classification
Chromatographic techniques are typically named by...
Gas Chromatography: Sample Injection Systems
Two primary injection methods are used...