Head and Neck Radiation Dose and Radiation Safety for Interventional Physicians
Kenneth Fetterly1, Beth Schueler2, Michael Grams3
1Department of Cardiovascular Diseases, Mayo Clinic, Rochester, Minnesota.
Objectives:
The first aim of this study was to assess the magnitude of radiation dose to tissues of the head and neck of physicians performing x-ray-guided interventional procedures. The second aim was to assess protection of tissues of the head offered by select wearable radiation safety devices.
Background:
Radiation dose to tissues of the head and neck is of significant interest to practicing interventional physicians. However, methods to estimate radiation dose are not generally available, and furthermore, some of the available research relating to protection of these tissues is misleading.
Methods:
Using a single representative geometry, scatter radiation dose to a humanoid phantom was measured using radiochromic film and normalized by the radiation dose to the left collar of the radioprotective thorax apron. Radiation protection offered by leaded glasses and by a radioabsorbent surgical cap was measured.
Results:
In the test geometry, average radiation doses to the unprotected brain, carotid arteries, and ocular lenses were 8.4%, 17%, and 50% of the dose measured at the left collar, respectively. Two representative types of leaded glasses reduced dose to the ocular lens on the side of the physician from which the scatter originates by 27% to 62% but offered no protection to the contralateral eye. The radioabsorbent surgical cap reduced brain dose by only 3.3%.
Conclusions:
A method by which interventional physicians can estimate dose to head and neck tissues on the basis of their personal dosimeter readings is described. Radiation protection of the ocular lenses by leaded glasses may be incomplete, and protection of the brain by a radioabsorbent surgical cap was minimal.
Related Concept Videos
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT