Whole blood selenium determination by inductively coupled plasma mass spectrometry
Dustin R Bunch1, Wendy Cieslak1, Sihe Wang1
1Department of Laboratory Medicine, Cleveland Clinic, Cleveland, OH, USA.
Objective:
To develop a sensitive method for accurately measuring whole blood selenium and determining an appropriate reference interval for the local Cleveland population.
Design And Methods:
The assay was developed and validated on an inductively coupled plasma mass spectrometry (ICP-MS) with a collision cell. Whole blood trace element free EDTA tubes were used to collect samples for the reference interval study (n=50). Samples were collected after at least 8h fast from healthy adults (76% females) with ages between 19 and 64yr. Whole blood aliquots (1mL) in acid washed cryogenic vials were stored at -70°C until analysis.
Results:
The method passed the matrix effect, interference (except for Gd), and carryover tests. The method had a linear range of 0.2-7.1μmol/L with accuracies of 87.1-118.1%. The total assay imprecision (CV) was <2.5% across the concentration levels tested. Comparison to another ICP-MS assay offered by an independent clinical lab yielded a Deming regression with a slope of 0.98, an intercept of 0.1μmol/L, a standard error of estimate of 0.1μmol/L, a correlation coefficient of 0.9846, and an average difference of 0.8%. The whole blood Se reference interval using a transformed parametric method was 2.2-3.5μmol/L.
Conclusions:
This whole blood Se ICP-MS methodology is sensitive and acceptable for patient testing.
More Related Videos
04:48Setup of Capillary Electrophoresis-Inductively Coupled Plasma Mass Spectrometry CE-ICP-MS for Quantification of Iron Redox Species FeII, FeIII
Published on: May 4, 2020
11:18A Practical Guide on Coupling a Scanning Mobility Sizer and Inductively Coupled Plasma Mass Spectrometer SMPS-ICPMS
Published on: July 11, 2017
Related Concept Videos
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Atomic Emission Spectroscopy: Lab
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...