Related Experiment Video
Updated: Mar 6, 2026

Home-Based Monitor for Gait and Activity Analysis
Published on: August 8, 2019
Validation of a Noninvasive, Disposable Activity Monitor for Clinical Applications
Background:
The study evaluated the concurrent and criterion validity of a new, disposable activity monitor designed to provide objective data on physical activity and energy expenditure in clinical populations.
Methods:
A sample of healthy adults (n = 52) wore the disposable Metria IH1 along with the established Sensewear armband (SWA) monitor for a 1-week period. Concurrent validity was examined by evaluating the statistical equivalence of estimates from the Metria and the SWA. Criterion validity was examined by comparing the relative accuracy of the Metria IH1 and the SWA for assessing walking/running. The absolute validity of the 2 monitors was compared by computing correlations and mean absolute percent error (MAPE) relative to criterion data from a portable metabolic analyzer.
Results:
The output from 2 monitors was highly correlated (correlations > 0.90) and the summary measures yielded nearly identical allocations of time spent in physical activity and energy expenditure. The monitors yielded statistically equivalent estimates and had similar absolute validity relative to the criterion measure (12% to 15% error).
Conclusions:
The disposable nature of the adhesive Metria IH1 monitor offers promise for clinical evaluation of physical activity behavior in patients. Additional research is needed to test utility for counseling and behavior applications.
More Related Videos
05:51Assessing the Accuracy of Fitness Smartwatch Data for Cardiovascular and Physical Activity Monitoring: A Validation Study in Digital Health
Published on: February 21, 2025
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
Related Concept Videos
Holter Monitor: 24-Hour Monitoring
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...