Validation of a modern second-check dosimetry system using a novel verification phantom.
Daniel G McDonald1, Dustin J Jacqmin1, Christopher J Mart1
1Department of Radiation Oncology, Medical University of South Carolina, Charleston, SC, USA.
The Mobius second-check dosimetry system accurately calculates radiation dose, agreeing with measurements within 3.5%. This system provides a reliable independent verification for modern treatment planning systems (TPS).
Area of Science:
- Medical Physics
- Radiation Oncology
- Dosimetry
Background:
- Accurate radiation dose calculation is critical in modern cancer treatment.
- Independent verification of treatment planning systems (TPS) enhances patient safety.
- The Mobius second-check dosimetry system offers an independent dose calculation and verification solution.
Purpose of the Study:
- To evaluate the Mobius second-check dosimetry system's accuracy by comparing its calculations to ionization-chamber measurements.
- To compare Mobius calculations against the Varian Eclipse treatment planning system (TPS) with the AcurosXB dose algorithm.
- To assess the clinical performance of the Mobius system by comparing patient doses calculated by Mobius and Eclipse.
Main Methods:
- Seventeen intensity-modulated clinical treatment plans were analyzed.
- Dose was recalculated using Mobius and Eclipse on the Mobius Verification Phantom™ (MVP).
- Calculated doses were compared to measurements from an A1SL ionization chamber in the MVP. Gamma analysis and dose metrics (target mean dose, D95%) were used for comparison.
Main Results:
- Mobius dose calculations agreed with measurements with an average difference of 0.3 ± 1.3%.
- Varian Eclipse calculations showed an average difference of -1.2 ± 0.7% compared to measurements.
- For pre-clinical cases, Mobius and Eclipse target mean dose differed by 1.0 ± 1.1%, with an average global gamma pass rate of 94.4% (3%, 2mm criteria).
Conclusions:
- The Mobius system accurately calculates dose, agreeing with measurements within 3.5% across all tested intensity-modulated plans.
- The Mobius system's dose calculation accuracy and independence make it suitable for rigorous second-check verification of TPS.
- Clinical implementation results were consistent with pre-clinical findings, demonstrating the system's reliability in practice.