Reconstruction of uptake patterns in PET: The influence of regularizing prior
Elisabetta De Bernardi1,2, Federico Fallanca2, Luigi Gianolli2
1Department of Medicine and Surgery, University of Milano-Bicocca, 20900, Monza, Italy.
Purpose:
The effects of regularizing priors on the maximum likelihood (ML) reconstruction of activity patterns in Positron Emission Tomography (PET) were assessed.
Methods:
Two edge-preserving priors (one originally proposed by Nuyts et al. and nowadays implemented and commercialized by General Electric Medical Systems as Q.Clear software, and a second one originally proposed by Rapisarda et al. and our group) were assessed and compared to a standard Ordered Subset (OS)-ML reconstruction, assumed as reference. The main difference between the two priors is that Nuyts prior (NY-p) penalizes relative voxel differences while Rapisarda prior (RP-p) absolute ones. Prior parameters were selected by imposing a reference noise texture inside uniform regions with activity comparable to that measured in 18 F-FluoroDeoxyGlucose (FDG) patient livers overall the field of view. Comparisons were then made: (a) on phantom data in terms of sphere recovery coefficients, ability to correctly reconstruct uniform irregularly shaped objects and heterogeneous patterns in patient backgrounds; (b) on patient data in terms of lesion detectability and image quality.
Results:
On phantoms, both priors succeeded in improving all the assessed features with respect to standard OS-ML reconstruction, mainly thanks to the better signal convergence and to the noise breakup control. On 10 mm spheres, an average recovery coefficient augment of 9% (NY-p) and 34% (RP-p) was obtained; homogeneity of uniform activity objects augmented of 4% (NY-p) and 11% (RP-p); accuracy in reconstructing heterogeneous lesions improved on average of 5% (NY-p) and 15% (RP-p). On patients, lesion detectability resulted improved (on 27 of 30 lesions), regardless of lesion anatomical districts and position in the scanner field of view. NY-p provides a spatial resolution and a noise texture more uniform in the field of view and an image quality similar to standard OS-ML. RP-p has instead a behavior more dependent on the local counting statistics that imposes a trade-off between spatial resolution uniformity and noise texture homogeneity.
Conclusions:
The assessed regularizing priors improve PET uptake pattern reconstruction accuracy. Therefore, they should be considered both for oncological lesion detection and uptake spatial distribution assessment. Pitfalls and open challenges are also discussed.
More Related Videos
11:09High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals
Published on: December 16, 2022
07:28Studying Metabolic Brain Connectivity Using 2-Deoxy-2-[18F]Fluoro-D-Glucose Dynamic Positron Emission Tomography at the Single-subject Level
Published on: January 24, 2025
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...