Premature Trigger of ERI in Medtronic EnRhythm Devices
Melissa E Middeldorp1, Rajiv Mahajan1, Adrian D Elliott1
1Centre for Heart Rhythm Disorders (CHRD), South Australian Health and Medical Research Institute (SAHMRI), University of Adelaide and Royal Adelaide Hospital, Adelaide, Australia.
Background:
Medical technology has made significant advances over the last few decades with smaller and more dynamic pacemakers. However, technical failures leading to premature replacement is a cause of concern. We present a series of Medtronic EnRhythm devices that reached premature elective replacement indicator (ERI).
Methods:
The database of Centre of Heart Rhythm Disorders was searched for EnRhythm device implantation from 2006 to 2011. Battery depletion <8.5 years was considered premature considering the projected average longevity to be 8.5-10.5 years. An unexpected premature ERI was defined when it was reached within 3 months of last normal check. Device follow-up was conducted every 3 months after advisory.
Results:
A total of 88 EnRhythm pacemakers were implanted. Over a median follow-up of 6.2 years (range: 0.3-9.2), 39 (44.3%) EnRhythm devices reached premature ERI. In 11 (28%), ERI was not recognized and patients were being investigated for other causes of unsteadiness or dyspnea prior to device check. Notably, three (7%) patients had premature ERI < 3.5 years. Ten (25.6%) had sudden and unexpected premature ERI. While asynchronous pacing was observed, there were no cases of absence of pacing.
Conclusions:
The rate of premature ERI for EnRhythm devices was 44.3%, significantly higher than reported by the manufacturer. Of concern, a sizeable proportion occurred unexpectedly, warranting more frequent reviews and empirical replacement in some patients. With the experience of the EnRhythm, appropriate monitoring strategies are recommended for future advisories.
Related Concept Videos
Dysrhythmias V: Evaluating Dysrhythmias
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Mechanism of Cardiac Arrhythmias
Cardiac Action Potential
The cardiac action potential process involves a series of phases characterized by the movement of ions across the cardiac cell membranes, leading to the depolarization and repolarization of the cardiac myocytes.
Ionic Basis of Cardiac Action Potentials
Dysrhythmias IV: Characteristics of Bradyarrhythmias
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...