An Innovation in Pacemaker Lead Implantation via Persistent Left Superior Vena Cava: The "3D Alpha Curve" Stylet
Jahangir Rashid Beig1, Mohd Iqbal Dar1, Nisar A Tramboo1
1Department of Cardiology, SKIMS, Srinagar, Jammu and Kashmir, India.
Abstract:
Persistent left superior vena cava (PLSVC) draining into coronary sinus is typically detected incidentally during transcatheter interventions using left subclavian venous approach. In our experience, we have encountered this anomaly on a few occasions and in all these cases we successfully implanted leads in the right ventricle (RV) by shaping the stylet into a "U-shaped" or "pigtail-shaped" curve. Herein, we report a case of an adult male who underwent successful dual-chamber pacemaker implantation via PLSVC through left axillary venous approach. In this case, we were unable to deliver the lead into RV using aforementioned stylets. As an innovation, we used a "three-dimensional alpha curve"-shaped stylet that facilitated an easy entry of pacing lead into the RV.
Related Concept Videos
Cardiac Catheterization III: Left Heart Catheterization
Cardiomyopathy V: Interprofessional Care
Cardiac Catheterization II: Right Heart Catheterization