Related Experiment Video
Updated: Mar 6, 2026

Author Spotlight: Insights into the Use of Apple-Derived Cellulose Scaffolds for Bone Tissue Engineering
Published on: February 23, 2024
Fracture with the newer bioresorbable vascular scaffolds
Mostafa Elwany1,2, Gaetano Di Palma1, Bernardo Cortese1,3
1Interventional Cardiology, A.O. Fatebenefratelli, Milano, Italy.
Abstract:
Bioresorbable vascular scaffolds (BVS) are being increasingly used in complex, real-world lesions. The possibility of strut fractures is a rare entity with the use of drug eluting stents (DES), but has not yet been adequately described for the BVS technology. In this report, we present a case of scaffold fracture of DEsolve BVS (Elixir Medical Corporation, USA) that was diagnosed both angiographically and by optical coherence tomography(OCT). Scaffold fracture with the newer BVS is a new Achilles' heel in the BVS technology. This gap in knowledge about the etiology such as over-expansion and potential predictors of strut fracture need to be deeply investigated before a widespread use of such technologies is adopted. © 2017 Wiley Periodicals, Inc.