MRI geometric distortion: Impact on tangential whole-breast IMRT
Amy Walker1,2,3, Peter Metcalfe1,2,3, Gary Liney1,2,3,4
1Centre for Medical Radiation Physics, University of Wollongong, Wollongong, NSW, Australia.
Journal of Applied Clinical Medical Physics
|March 16, 2017
Summary
Magnetic resonance imaging (MRI) geometric distortions can impact whole-breast intensity-modulated radiotherapy (IMRT) planning. While system distortions alone had minimal effect, combined MRI system and patient distortions led to clinically unacceptable plans in some cases.
Area of Science:
- Medical Physics
- Radiotherapy
- Medical Imaging
Background:
- Accurate target delineation is crucial for effective radiotherapy.
- Magnetic resonance imaging (MRI) offers superior soft-tissue contrast for treatment planning.
- Geometric distortions in MRI can potentially compromise treatment accuracy.
Purpose of the Study:
- To evaluate the impact of MRI geometric distortions on whole-breast intensity-modulated radiotherapy (IMRT) planning.
- To assess the influence of residual system and patient-induced distortions on treatment dosimetry.
- To determine the clinical acceptability of MRI-based radiotherapy plans.
Main Methods:
- Retrospective analysis of 18 patients undergoing whole-breast radiotherapy with both CT and MRI data.
- Imaging distortion phantoms on 3T and 1T MRI systems to quantify residual system distortions.
- Simulating MRI-based planning by applying distortion maps to CT datasets.
- Generating and evaluating IMRT plans based on distorted and original CT data.
Main Results:
- Maximum residual systematic distortions measured were 7.9 mm (3T) and 11.9 mm (1T).
- Residual system distortions alone did not lead to clinically unacceptable plans.
- Combined MRI systematic and patient-induced distortions resulted in unacceptable dosimetry for 8 plans.
Conclusions:
- Residual MRI system distortions alone have a minimal impact on whole-breast IMRT dosimetry.
- The combination of MRI systematic and patient-related distortions can compromise treatment plan quality.
- Careful consideration of MRI distortions is necessary for MRI-only radiotherapy planning.