Background And Aims:
The colonic self-expanding metal stent (SEMS) has been widely used for "bridge to surgery" and palliative therapy. However, if the spread of SEMS is insufficient, not only can a decompression effect not be obtained but also perforation and obstructive colitis can occur. The mechanism of occurrence of obstructive colitis and perforation was investigated by flow dynamics.
Methods:
Bernoulli's principle was applied, assuming that the cause of inflammation and perforation represented the pressure difference in the proximal lumen and stent. The variables considered were proximal lumen diameter, stent lumen diameter, flow rate into the proximal lumen, and fluid density. To model the right colon, the proximal lumen diameter was set at 50 mm. To model the left-side colon, the proximal lumen diameter was set at 30 mm.
Results:
For both the right colon model and the left-side colon model, the difference in pressure between the proximal lumen and the stent was less than 20 mmHg, when the diameter of the stent lumen was 14 mm or more. Both the right colon model and the left-side colon model were 30 mmHg or more at 200 mL s-1 when the stent lumen was 10 mm or less. Even with an inflow rate of 90-110 mL s-1, the pressure was 140 mmHg when the stent lumen diameter was 5 mm.
Conclusion:
In theory, in order to maintain the effectiveness of SEMS, it is necessary to keep the diameter of the stent lumen at 14 mm or more.
More Related Videos
11:00Experimental Investigation of Secondary Flow Structures Downstream of a Model Type IV Stent Failure in a 180° Curved Artery Test Section
Published on: July 19, 2016
09:40An Ivor Lewis Esophagectomy Designed to Minimize Anastomotic Complications and Optimize Conduit Function
Published on: April 17, 2020
Related Concept Videos
Bernoulli's Principle: Applications
Entrainment devices use a high fluid speed to create low pressures and, thus, entrain one fluid into another. Some examples of these devices are given below:
Bernoulli's Principle
Bernoulli's principle has several...
Bernoulli's Equation for Flow Normal to a Streamline
The pressure difference depends on the fluid's velocity and radius of curvature. The pressure variation is minimal in flows with nearly straight streamlines. However, the...
Bernoulli's Equation for Flow Along a Streamline
Bernoulli's Equation: Problem Solving
The first step is to compute the cross-sectional areas of the pipe and the Venturi throat to analyze the pressure difference indicated by the pressure gauge. Next, the continuity equation is...
Bernoulli's Equation