Related Experiment Video
Updated: Mar 6, 2026

12:00
Evaluation of a Novel Laser-assisted Coronary Anastomotic Connector - the Trinity Clip - in a Porcine Off-pump Bypass Model
Published on: November 24, 2014
15.9K
The second-generation eCLIPs Endovascular Clip System: initial experience
Albert H Chiu1,2,3, Joost De Vries4, Cian J O'Kelly5
11Neurovascular Program.
Journal of Neurosurgery
|March 18, 2017
Summary
The second-generation eCLIPs device offers a viable treatment for wide-necked bifurcation aneurysms, showing comparable occlusion rates to other devices. This endovascular clip system demonstrates effectiveness in complex intracranial aneurysm treatment.
Area of Science:
- Neurosurgery
- Interventional Neuroradiology
- Medical Devices
Background:
- Treatment of wide-necked intracranial aneurysms, especially at bifurcations, presents challenges with existing methods.
- High recanalization and complication rates are associated with current treatments for these complex aneurysms.
Purpose of the Study:
- To evaluate the evolution, safety, and efficacy of the second-generation eCLIPs (Endovascular Clip System) device for treating bifurcation aneurysms.
- To assess the device's performance in terms of deployment, complication rates, and long-term angiographic and clinical outcomes.
Main Methods:
- A registry study involving 13 international centers treated patients with the second-generation eCLIPs device.
- Data collected included aneurysm characteristics, procedural details, device coverage, coil retention, and patient follow-up (imaging and clinical).
- Follow-up assessed angiographic results using the modified Raymond-Roy Classification and need for re-treatment.
Main Results:
- Successful eCLIPs device placement was achieved in 76% of 33 treated patients, with a learning curve observed in early deployment.
- Low rates of periprocedural complications were noted, including transient ischemic attacks and asymptomatic thrombotic events.
- At an average 8-month follow-up, 81% of patients achieved a modified Raymond-Roy Classification of I or II, with 42.9% showing improvement in Raymond grade.
Conclusions:
- The second-generation eCLIPs device is a feasible treatment option for intracranial bifurcation aneurysms.
- Aneurysm occlusion rates achieved with eCLIPs are comparable to other bifurcation support devices in initial clinical series.
- The device shows promise for managing complex wide-necked aneurysms, particularly at bifurcation sites.
Keywords:
ICA = internal carotid arteryMCA = middle cerebral arteryMRRC = modified Raymond-Roy ClassificationPCA = posterior cerebral arteryWEB = Woven EndoBridge Aneurysm Embolization Systembifurcation aneurysmembolizationflow diversioninterventional neuroradiologyinterventional neurosurgeryintracranial aneurysmradiologytherapeuticvascular disorders