Related Experiment Video
Updated: Mar 6, 2026

Remote Magnetic Navigation for Accurate, Real-time Catheter Positioning and Ablation in Cardiac Electrophysiology Procedures
Published on: April 21, 2013
Slow Pathway Radiofrequency Ablation Using Magnetic Navigation: A Description of Technique and Retrospective Case
Abhishek Bhaskaran1, Maha Albarri1, Neil Ross2
1Cardiology Department, Westmead Hospital, Sydney, NSW, Australia; Sydney Medical School, University of Sydney, Sydney, NSW, Australia.
Background:
The Magnetic Navigation System (MNS) catheter was shown to be stable in the presence of significant cardiac wall motion and delivered more effective lesions compared to manual control. This stability could potentially make AV junctional re-entrant tachycardia (AVNRT) ablation safer. The aim of this study is to describe the method of mapping and ablation of AVNRT with MNS and 3-D electro-anatomical mapping system (CARTO, Biosense Webster, Diamond bar, CA, USA) anatomical mapping, with a view to improve the safety of ablation.
Methods:
The method of precise mapping and ablation with MNS is described. Consecutive AVNRT cases (n=30) from 2012 January to 2015 November, in which magnetic navigation was used, are analysed.
Results:
Ablation was successful in 27 (90%) out of 30 patients. In three cases, ablation was abandoned due to the proximity of the three-dimensional His image to the potential ablation site. No complications, including AV nodal injury, occurred. The distance from the nearest His position to successful ablation site in both LAO and RAO projections of CARTO images was 26.4±8.8 and 27±7.7mm respectively. Only in two (9%) patients, ablation needed to be extended superior to the plane of coronary sinus ostium, towards the His bundle region, to achieve slow pathway modification.
Conclusion:
AVNRT ablation with MNS allows for accurate mapping of the AV node and stable ablation at a safe distance, which could help avoid AV nodal injury. We recommend this modality for younger patients with AVNRT.
More Related Videos
10:46Non-fluoroscopic Catheter Tracking for Fluoroscopy Reduction in Interventional Electrophysiology
Published on: May 26, 2015
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017