Insertion, Maintenance, and Removal of the Percutaneous Dual Lumen Cannula Right Ventricular Assist Device
Published on: July 20, 2022
A Numerical Simulation Comparing a Cavopulmonary Assist Device and VA ECMO for Failing Fontan Support
Po-Lin Hsu1, Dongfang Wang, Cherry Ballard-Croft
1From the *Artificial Organ Technology Laboratory, Biomanufacturing Centre, School of Mechanical and Electrical Engineering, Soochow University, Suzhou, China; and †Department of Surgery, University of Kentucky College of Medicine, Lexington, Kentucky.
Abstract:
A cavopulmonary assist device (CPAD) has been developed for failing Fontan support. This CPAD pumps blood from superior/inferior vena cavae (SVC/IVC) to pulmonary artery. In this study, we compared failing Fontan support with CPAD versus veno-arterial extracorporeal membrane oxygenation (VA ECMO) in silico. A numerical lumped-parameter circulatory system model was used to simulate Fontan circulation. Failing Fontan was simulated by increased pulmonary resistance. Central venous pressure (CVP), mean pulmonary arterial pressure (mPAP), left atrial pressure (LAP), and univentricular outflow (CO) were simulated and compared with published clinical data. The CPAD and VA ECMO were simulated using 1-5 L/min pump flows. In agreement with published clinical data, the simulated failing Fontan condition had increased CVP (19 mmHg) and mPAP (18 mmHg) with decreased LAP (7 mmHg) and cardiac output (3 L/min) compared with functional Fontan condition. The CPAD achieved total Fontan assistance with pump flows higher than original CO. Veno-arterial extracorporeal membrane oxygenation provided partial Fontan assistance with low pump flows. Blood went through pulmonary circulation with CPAD whereas VA ECMO bypassed pulmonary circulation and diminished univentricular blood flow (0.8 L/min). This in silico study demonstrated that CPAD preserved heart/lung function whereas VA ECMO had very low univentricular flow, potentially leading to thrombosis or univentricular atrophy.
Related Concept Videos
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...