Thoracic electrical bioimpedance versus suprasternal Doppler in emergency care
Mohammed H Elwan1,2, Jeremy Hue3, Samira J Green4
1Department of Emergency Medicine, Alexandria University, Alexandria, Egypt.
Non-invasive cardiac output (CO) monitoring using suprasternal Doppler and thoracic electrical bioimpedance showed differences in measurements. These methods are not interchangeable for emergency care, highlighting the need for careful selection of monitoring devices.
Area of Science:
- Critical care medicine
- Cardiovascular physiology
- Medical device technology
Background:
- Cardiac output (CO) monitoring is crucial in emergency departments (EDs).
- Thoracic electrical bioimpedance and suprasternal Doppler are promising non-invasive CO monitoring methods.
- Direct comparisons between these two methods in emergency settings are lacking.
Purpose of the Study:
- To assess the feasibility of using suprasternal Doppler and bioimpedance for CO monitoring in emergency care.
- To compare hemodynamic data obtained from both suprasternal Doppler and bioimpedance monitors.
- To evaluate the interchangeability of these non-invasive CO monitoring techniques.
Main Methods:
- Hemodynamic measurements were performed on the same patient cohort.
- Utilized bioimpedance (Niccomo, Medis, Germany) and suprasternal Doppler (USCOM, Sydney, Australia) for data acquisition.
- Assessed the usability and comparability of data from both devices.
Main Results:
- Supratemporal Doppler yielded usable CO data in 97% of patients, while bioimpedance provided usable data in 87%.
- Median CO measurements were 3.4 L/min lower with suprasternal Doppler compared to bioimpedance.
- Median stroke volume was 51 mL lower when measured by suprasternal Doppler.
Conclusions:
- The two non-invasive cardiac monitoring methods, suprasternal Doppler and bioimpedance, are not interchangeable.
- The selection of an appropriate non-invasive cardiac monitor is significant in emergency care.
- Further research is needed to establish clear criteria for choosing between these monitoring technologies.
Related Concept Videos
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Acute Coronary Syndrome III: Diagnostic Studies
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for...
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...