Accuracy of PDFF estimation by magnitude-based and complex-based MRI in children with MR spectroscopy as a reference
William M Haufe1, Tanya Wolfson2, Catherine A Hooker1
1Liver Imaging Group, Department of Radiology, University of California - San Diego, San Diego, California, USA.
Purpose:
To assess and compare the accuracy of magnitude-based magnetic resonance imaging (MRI-M) and complex-based MRI (MRI-C) for estimating hepatic proton density fat fraction (PDFF) in children, using MR spectroscopy (MRS) as the reference standard. A secondary aim was to assess the agreement between MRI-M and MRI-C.
Materials And Methods:
This was a HIPAA-compliant, retrospective analysis of data collected in children enrolled in prospective, Institutional Review Board (IRB)-approved studies between 2012 and 2014. Informed consent was obtained from 200 children (ages 8-19 years) who subsequently underwent 3T MR exams that included MRI-M, MRI-C, and T1 -independent, T2 -corrected, single-voxel stimulated echo acquisition mode (STEAM) MRS. Both MRI methods acquired six echoes at low flip angles. T2*-corrected PDFF parametric maps were generated. PDFF values were recorded from regions of interest (ROIs) drawn on the maps in each of the nine Couinaud segments and three ROIs colocalized to the MRS voxel location. Regression analyses assessing agreement with MRS were performed to evaluate the accuracy of each MRI method, and Bland-Altman and intraclass correlation coefficient (ICC) analyses were performed to assess agreement between the MRI methods.
Results:
MRI-M and MRI-C PDFF were accurate relative to the colocalized MRS reference standard, with regression intercepts of 0.63% and -0.07%, slopes of 0.998 and 0.975, and proportion-of-explained-variance values (R2 ) of 0.982 and 0.979, respectively. For individual Couinaud segments and for the whole liver averages, Bland-Altman biases between MRI-M and MRI-C were small (ranging from 0.04 to 1.11%) and ICCs were high (≥0.978).
Conclusion:
Both MRI-M and MRI-C accurately estimated hepatic PDFF in children, and high intermethod agreement was observed.
Level Of Evidence:
1 Technical Efficacy: Stage 2 J. Magn. Reson. Imaging 2017;46:1641-1647.
More Related Videos
13:56The Use of Magnetic Resonance Spectroscopy as a Tool for the Measurement of Bi-hemispheric Transcranial Electric Stimulation Effects on Primary Motor Cortex Metabolism
Published on: November 19, 2014
07:59Fat-Water Phantoms for Magnetic Resonance Imaging Validation: A Flexible and Scalable Protocol
Published on: September 7, 2018
Related Concept Videos
Magnetic Resonance Imaging
NMR Spectrometers: Resolution and Error Correction
¹H NMR: Interpreting Distorted and Overlapping Signals
As Δν decreases and the signals move closer, the doublets appear increasingly distorted. The intensities of the inner lines increase at the cost of those of the outer lines as the signals are...