Validation of the G.LAB MD2231 digital automatic wrist blood pressure monitor according to multiple protocols
Yinghui Jiao1, Qingfen Guan, Lili Wu
1aDepartment of Neurology bPrenatal Diagnosis Laboratory, Weifang People's Hospital, Weifang cDepartment of Cardiology, Mizhou Health Center, Zhucheng City, Shandong, China.
Objective:
The objective of this study was to validate the G.LAB MD2231 digital automatic wrist BP monitor according to multiple protocols, which included the European Society of Hypertension International Protocol (ESH-IP) revision 2010, the British Hypertension Society (BHS), and the Association for the Advancement of Medical Instrumentation (AAMI)/the International Organization for Standardization (ISO) protocols.
Participants And Methods:
A total of 33 participants were initially included for validation on the basis of the ESH-IP revision 2010, and then the device was further assessed on 85 participants according to the BHS and AAMI/ISO protocols. The validation procedures of the protocols were followed precisely.
Results:
The G.LAB MD2231 achieved the requirements of the ESH-IP revision 2010 with mean differences of 1.99±5.13 and -0.85±4.83 mmHg for systolic and diastolic BP, respectively. Furthermore, the A/A grade was obtained according to the BHS protocol with mean differences of 2.19±6.89 and -1.03±6.56 mmHg for systolic and diastolic BP, respectively, and the two criteria of the AAMI/ISO were also fulfilled.
Conclusion:
The G.LAB MD2231 digital automatic wrist BP monitor fulfilled the ESH-IP revision 2010, the BHS, and the AAMI/ISO protocols, which can be recommended for BP measurement in the general population.
More Related Videos
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
05:01A Detailed Protocol for Physiological Parameters Acquisition and Analysis in Neurosurgical Critical Patients
Published on: October 17, 2017
Related Concept Videos
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Pre-Procedural Guidelines for Assessing Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Errors occurring during blood pressure monitoring
Several factors...