Standardized Technique of Aortic Valve Re-implantation for Valve-sparing Aortic Root Replacement
Published on: December 11, 2017
14.8K
Transcatheter Aortic Valve Replacement With Different Valve Types in Elliptic Aortic Annuli
Yoshio Maeno1, Yigal Abramowitz1, Sung-Han Yoon1
1Cedars-Sinai Medical Center, Heart Institute.
Summary
Elliptic annuli in transcatheter aortic valve replacement (TAVR) impact self-expanding (SE) valve success. Optimal oversizing of SE valves for elliptic annuli significantly improves device success, achieving rates similar to balloon-expandable (BE) valves.
Area of Science:
- Cardiovascular Medicine
- Interventional Cardiology
- Medical Devices
Background:
- Transcatheter aortic valve replacement (TAVR) is a key treatment for aortic stenosis.
- Device success rates can be influenced by patient-specific aortic annulus anatomy.
- The impact of elliptic annuli on TAVR device success requires further investigation.
Purpose of the Study:
- To evaluate the influence of aortic annulus ellipticity on acute device success in self-expanding (SE) versus balloon-expandable (BE) TAVR.
- To compare device success rates between SE-TAVR and BE-TAVR in patients with circular versus elliptic aortic annuli.
Main Methods:
- Aortic annulus ratio (AAR) was calculated (short axis diameter/long axis diameter) to define circular (AAR ≥0.82) and elliptic (AAR <0.82) annuli.
- Device success was assessed using Valve Academic Research Consortium-2 definitions.
- Outcomes for SE-TAVR and BE-TAVR were compared within each annulus group, with further analysis of oversizing effects.
Main Results:
- For circular annuli, both SE and BE valves demonstrated high device success rates (90.5% vs. 95.0%, P=0.14).
- For elliptic annuli, SE valves had significantly lower device success rates than BE valves (82.5% vs. 95.3%, P=0.002).
- Optimal oversizing (≥17.5%) of SE valves for elliptic annuli led to a significant increase in device success, achieving rates comparable to BE valves (96.2% vs. 95.3%).
Conclusions:
- Aortic annulus ellipticity significantly affects SE-TAVR device success.
- While SE-TAVR shows lower success in elliptic annuli, optimal oversizing mitigates this disadvantage.
- Both SE and BE TAVR achieve high device success in elliptic annuli when SE valves are optimally oversized.