Prospective microbiologic evaluation of the forceps elevator in closed-channel duodenoscopes after reprocessing
Helga Paula1, Barbara Tribl2, Elisabeth Presterl1
1Department of Hospital Hygiene and Infection Control, Vienna General Hospital, Medical University of Vienna, Vienna, Austria.
Diligent reprocessing of closed-channel duodenoscopes, like the TJF-Q180V model, effectively prevents bacterial contamination. This study found no microorganisms after using validated reprocessing standards, ensuring patient safety in healthcare facilities.
Area of Science:
- Infection control in healthcare settings
- Medical device reprocessing efficacy
- Gastroenterology and Hepatology
Background:
- Endoscopes are known vectors for bacterial transmission in healthcare.
- Multidrug-resistant infections linked to reprocessed duodenoscopes, particularly after endoscopic retrograde cholangiopancreatography.
- Difficulties in reprocessing duodenoscopes, especially the forceps elevator, contribute to contamination.
Purpose of the Study:
- To prospectively evaluate the efficacy and safety of current reprocessing procedures for Olympus TJF-Q180V duodenoscopes.
- To address concerns raised by the FDA regarding closed-channel duodenoscope reprocessing.
- To assess the risk of bacterial contamination in a clinical setting.
Main Methods:
- Prospective collection of microbiologic surveillance samples from 6 TJF-Q180V duodenoscopes.
- Sampling occurred from August 2015 to March 2016 at the Division of Gastroenterology and Hepatology.
- ESwab collection system used for all samples.
Main Results:
- A total of 237 microbiologic samples were collected from the forceps elevator.
- No microorganism growth was detected in any of the collected samples.
- This indicates successful bacterial removal with the current reprocessing protocol.
Conclusions:
- Diligent adherence to validated reprocessing standards, per manufacturer recommendations, can ensure the safe use of closed-channel endoscopes.
- Further development of validated adaptations for closed-channel duodenoscopes is necessary for simpler and safer reprocessing.
- Establishment of comprehensive postmarket surveillance is crucial for ongoing safety assurance.
Related Concept Videos
Endoscopic Procedures V: ERCP
Patient...
Endoscopic Procedures I: Esophagogastroduodenoscopy
During an EGD, the endoscope can be used to:
Endoscopic Procedures III: Video Capsule Endoscopy
Endoscopic Procedures IV: Sigmoidoscopy and Laproscopy
Sigmoidoscopy
Sigmoidoscopy is a diagnostic procedure that uses a flexible sigmoidoscope equipped with a light source and camera to examine the rectum and sigmoid colon. The procedure involves inserting the tube through the anus...