Related Experiment Video
Updated: Mar 5, 2026

Utilizing Percutaneous Ventricular Assist Devices in Acute Myocardial Infarction Complicated by Cardiogenic Shock
Published on: June 12, 2021
Use of a parallel stiff wire to facilitate percutaneous Impella RP ventricular assist device positioning
Francesco Giannini1, Matteo Montorfano2, Richard J Jabbour2
1Interventional Cardiology Unit, San Raffaele Hospital, Milan, Italy; EMO-GVM Centro Cuore Columbus, Milan, Italy.
Abstract:
Despite optimal medical management, some patients with severe right ventricular failure fail to respond and may benefit from additional support with the implantation of a RV assist device. Experience to date with Impella RP is limited. We report a case of percutaneous Impella RP implantation, using a parallel stiff wire to reduce anatomical tortuosity by acting as a buddy-wire to facilitate device implantation and reduce the risk of tricuspid ring damage in a patient recently treated with tricuspid ring annuloplasty.