V Rigo1, C Debauche2, P Maton3
1Neonatology Division, CHU de Liège- CHR Citadelle, Liège, Belgium.
Aim:
Different catheters can be used for less invasive surfactant therapy (LIST): feeding tubes inserted with or without Magill forceps, different angiocatheters and centre specific devices, such as umbilical catheters affixed to a stylet. This study compared the effectiveness of LIST devices and endotracheal tubes (ETT).
Methods:
Video recordings of 20 neonatologists simulating different LIST techniques on two manikin heads were analysed. Procedural effectiveness was evaluated by the duration of procedures and failure rates. Ease of use was scored.
Results:
The median procedure time for the Neonatal Intubation Trainer was significantly longer with feeding tubes without Magill forceps. For the more difficult ALS Baby Trainer, successful procedures lasted a median of 24 (17-32) seconds with ETT, 24 (15-36) seconds with stylet-guided catheters and 34 (27-46) seconds and 37 (29-42) seconds with 13-cm and 30-cm angiocatheters, respectively. Both methods using feeding tubes were statistically slower than ETT intubation, lasting 32 (25-44) seconds and 39 (27-95) seconds with or without Magill forceps. Failure rates (7-20%) were no different between the LIST methods. Techniques using feeding tubes were rated as more difficult.
Conclusion:
Only rigid or stylet-guided catheters required tracheal catheterisation times similar to those of endotracheal intubation and neonatologists found them easier.
More Related Videos
06:22Surfactant Depletion Combined with Injurious Ventilation Results in a Reproducible Model of the Acute Respiratory Distress Syndrome ARDS
Published on: April 7, 2021
07:20Lavage-induced Surfactant Depletion in Pigs As a Model of the Acute Respiratory Distress Syndrome ARDS
Published on: September 7, 2016
Related Concept Videos
Suctioning the Nasopharyngeal Airway
Equipment Required
Tracheostomy Suctioning II: Procedure