Clinical transition to model-based dose calculation algorithm: A retrospective analysis of high-dose-rate tandem and
Dayee Jacob1, Melissa Lamberto2, Lana DeSouza Lawrence1
1Department of Radiation Oncology, Helen F Graham Cancer Center and Research Institute, Christiana Care Health System, Newark, DE.
Purpose:
To retrospectively compare clinical dosimetry of CT-based tandem-ring treatment plans using a model-based dose calculation algorithm (MBDCA) with the standard TG-43-based dose formalism.
Methods And Materials:
A cohort of 10 cervical cancer cohorts treated using the tandem and ring high-dose-rate applicators were evaluated. The original treatment plans were created using the department CT-based volume optimization clinical standards. All plans originally calculated with TG-43 dose calculation formalism were recalculated using the MBDCA algorithm. The gross target volume and organs at risk (OARs) were contoured on each data set along with significant heterogeneities like air in cavity and high-density plastic tandem and ring components. The patient tissue was modeled as homogenous liquid water. D90, D95, and D100 for gross target volume, D0.1cm, D1.0cm, and D2.0cm for bladder, rectum, and sigmoid were extracted from dose-volume histograms for TG-43 and MBDCA calculated plans. Mean absolute difference ± 2σ in the above metrics was calculated for each plan.
Results:
Using the manual applicator contouring method, MBDCA plans (n = 10) showed 2.1 ± 1.1% reduction in dose to Point A average, 2.6 ± 0.9% reduction in Target D90 dose, and 2.1 ± 0.3% dose reduction to OARs. Results from plans using vendor supplied solid applicator models (n = 5) showed 2.2 ± 1.10% reduction in dose to Point A average, 2.7 ± 0.2% reduction in Target D90 dose, and 2.7 ± 1.0% dose reduction on average to OARs.
Conclusion:
For unshielded plastic gynecologic applicators, minimal dosimetric changes (<5%) were found using MBDCA relative to standard TG-43. Use of solid applicator model is more efficient than manual applicator contouring and also yielded similar MBDCA dosimetric results. Currently, TG-186 dose calculations should be reported along TG-43 until we obtain studies with larger cohorts to fully realize the potential of MBDCA dosimetry.