Percutaneous Mitral Valve Repair With the MitraClip in Primary Compared With Secondary Mitral Valve Regurgitation
Julia Seeger, Patrick Müller, Birgid Gonska
1Department of Internal Medicine II, University of Ulm, Albert-Einstein-Allee 23, 89081 Ulm, Germany. jochen.woehrle@uniklinik-ulm.de.
Aims:
To compare early device success, procedural success, and 30-day safety endpoint according to the new Mitral Valve Academic Research Consortium criteria (MVARC) in severe primary and secondary mitral regurgitation (MR) patients.
Methods And Results:
A total of 210 patients were enrolled; 105 patients with primary MR were compared with 105 patients with secondary MR. All patients were highly symptomatic (New York Heart Association III/IV 79.0% vs 87.6%). Decision for MitraClip therapy was done by the heart team. Patients were on optimal medical heart failure therapy. Preprocedural MR grade was 3.4 ± 0.5 in secondary MR vs 3.7 ± 0.4 in primary MR (P<.001). Device success according to MVARC was high in both groups (93.3% in secondary MR vs 94.3% in primary MR), treated with 1.4 ± 0.6 vs 1.3 ± 0.5 MitraClips (P=.14). Reduction of New York Heart Association class from baseline to 30-day follow-up was 1.7 ± 1.1 in secondary MR vs 2.2 ± 1.2 in primary MR (P<.01). Rate of MVARC minor vascular complications was low. Thirty-day safety endpoint according to MVARC criteria was low in both groups (4.8% in secondary MR vs 5.7% in primary MR (P=non-significant).
Conclusion:
Percutaneous mitral valve repair using the MitraClip device is safe and effective in patients with primary and secondary MR, with a high early device success rate and low 30-day safety endpoint according to the MVARC criteria.