Reducing mass peak instability caused by the phase changes of RF and AC signals in a rectilinear ion-trap analyzer
Xinqiong Lu1, Kai Ni1, Quan Yu1
1Division of Advanced Manufacturing, Graduate School at Shenzhen, Tsinghua University, Shenzhen 518055, China.
Abstract:
For an ion trap with resonance ejection, peak intensity and peak position of the acquired mass spectra are affected by the phase difference between the radio frequency (RF) and auxiliary alternating current (AC) potentials. To ensure measurement stability, RF and AC phase-locking is commonly used in commercial ion trap mass spectrometers. In this study, a compact electronic control system was developed to accurately regulate the RF and AC phases and was employed in a photoionization rectilinear ion trap (RIT) mass spectrometer. We found that the phase-locking method was defective in multicomponent analysis because the optimal RF and AC phase difference was usually different for different m/z peaks. After studying and characterizing the relationship between the peaks and the RF and AC phases, a correction method based on data processing was used to improve the peaks' stability and accuracy. The results show that the fluctuations of both peak intensity and peak position were significantly reduced and that the instrument presented satisfying reproducibility and quantitative ability.
Related Concept Videos
Mass Spectrometers
Mass Spectrum: Interpretation
Mass Analyzers: Overview
Mass Analyzers: Common Types
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences