Related Experiment Video
Updated: Jul 15, 2026

Angle-resolved Photoemission Spectroscopy At Ultra-low Temperatures
Published on: October 9, 2012
Electron energy loss spectroscopy with parallel readout of energy and momentum
Harald Ibach1, François C Bocquet2, Jessica Sforzini2
1Peter Grünberg Institut (PGI-6), Forschungszentrum Jülich, 52425 Jülich, Germany.
Abstract:
We introduce a high energy resolution electron source that matches the requirements for parallel readout of energy and momentum of modern hemispherical electron energy analyzers. The system is designed as an add-on device to typical photoemission chambers. Due to the multiplex gain, a complete phonon dispersion of a Cu(111) surface was measured in 7 min with 4 meV energy resolution.
Related Concept Videos
Molecular Spectroscopy: Absorption and Emission
Atomic Absorption Spectroscopy: Instrumentation
The atomizer used in AAS can be either a flame atomizer or an...
Atomic Emission Spectroscopy: Overview
Atomic Emission Spectroscopy: Instrumentation
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.
Atomic Emission Spectroscopy: Lab