A Novel Anterior Transpedicular Screw Artificial Vertebral Body System for Lower Cervical Spine Fixation: A Finite
Weidong Wu1, Chun Chen2, Jinpei Ning3
1Department of Anatomy, Guangdong Provincial Medical Biomechanical Key Laboratory, Academy of Orthopedics of Guangdong Province, Southern Medical University, Guangzhou 510515, China;Wuhan Concrete Technology Company Limited, Gaoxin Avenue 818, Wuhan 430200, Hubei, China
Abstract:
A finite element model was used to compare the biomechanical properties of a novel anterior transpedicular screw artificial vertebral body system (AVBS) with a conventional anterior screw plate system (ASPS) for fixation in the lower cervical spine. A model of the intact cervical spine (C3-C7) was established. AVBS or ASPS constructs were implanted between C4 and C6. The models were loaded in three-dimensional (3D) motion. The Von Mises stress distribution in the internal fixators was evaluated, as well as the range of motion (ROM) and facet joint force. The models were generated and analyzed by mimics, geomagic studio, and ansys software. The intact model of the lower cervical spine consisted of 286,382 elements. The model was validated against previously reported cadaveric experimental data. In the ASPS model, stress was concentrated at the connection between the screw and plate and the connection between the titanium mesh and adjacent vertebral body. In the AVBS model, stress was evenly distributed. Compared to the intact cervical spine model, the ROM of the whole specimen after fixation with both constructs is decreased by approximately 3 deg. ROM of adjacent segments is increased by approximately 5 deg. Facet joint force of the ASPS and AVBS models was higher than those of the intact cervical spine model, especially in extension and lateral bending. AVBS fixation represents a novel reconstruction approach for the lower cervical spine. AVBS provides better stability and lower risk for internal fixator failure compared with traditional ASPS fixation.