A Word of Caution for Extra Large Self-expandable Nitinol Stents
Andrés Reyes Valdivia1, Africa Duque Santos1, Francisco Alvarez Marcos2
1Department of Vascular and Endovascular Surgery, Ramón y Cajaĺs University Hospital, Madrid, Spain.
Background:
To report an unsuccessful use of large nitinol stent to prevent proximal endoleak in endograft treatment for juxtarenal aortic aneurysm.
Case Report:
An 82-year-old male presented coronary disease and severe chronic obstructive pulmonary disease requiring oxygen support. A large 80-mm juxtarenal aneurysm was found on routine urological examination. Neck features were as follows: 7 mm and high anterior-posterior angulation. ASA risk score IV was given. Patient wanted to be treated. We decided a 1-shot intervention based on self-expandable suprarenal fixation endograft with adjunctive extra large self-expandable (high radial force) nitinol stent. Control computed tomography scan demonstrates nitinol stent severe infolding and a possible perforation of the primary endograft. At 6-month follow-up, sac reveals enlargement of 5 mm and 6% volume increase. Patient does not want further treatments due to high risk of procedures.
Conclusions:
Extreme caution should be taken when using self-expandable extra large nitinol stents for preventive matters. Oversizing should be carefully considered, not necessarily reaching the endograft chosen size. When short and angulated neck morphology co-exists, other devices or procedures may be a better option.